Һѹ»úÀ­ÉìÊÔÑé»ú:À­Á¦²âÊÔ»úÆ«ÒƳ£¼û¹ÊÕϼìÐÞ

ä¯ÀÀÁ¿:2103? À´Ô´:www.quanlitest.com??×÷Õß:AGÆì½¢Ìü¹ÙÍø²âÊÔ Ê±¼ä:2019-10-12

¡¡¡¡ÒºÑ¹ÍòÄÜÊÔÑé»ú:À­Á¦ÊÔÑé»úλÒÆ·ºÆð¹ÊÕϺóµÄάÐÞ°ì·¨£º

¡¡¡¡Hydraulic universal testing machine: Maintenance steps after displacement failure of tension testing machine:

¡¡¡¡1¡¢Ê×ÏȶÔλÒÆ´«¸ÐÆ÷½øÐмì²é£º

¡¡¡¡1. First, the displacement sensor is checked.

¡¡¡¡ÒòΪÀ­Á¦ÊÔÑé»úϵͳÊÇÒ»¸ö±Õ»·ÏµÍ³£¬ÒªÏëÖªµÀÊÇÄĸö²¿·Ö·ºÆðÁËÎÊÌ⣬¾ÍÒ»¶¨ÒªÇжÏϵͳÖеĻØ·£¬Ê¹Æä³ÉΪһ¸ö¿ª»·µÄϵͳ¡£´Ó¶ø¿ÉÒÔÅжÏÊÔÑé»úµÄ´«¸ÐÆ÷ÐźÅÊDz»ÊÇÕý³£¡£Õâ¸öʱºòÒ²Òª¶Ï¿ªÎ»ÒÆ´«¸ÐÆ÷µÄ·´ÏìÐźÅ£¬Ê¹Æä³ÉΪ¿ª»·¡£´ËʱÊÂÇéÕ¾·¢³öÒ»¸ö¿ØÖÆÐж¯Æ÷Òƶ¯µÄÐźÅ£¬¶Ô´«¸ÐÆ÷µÄ·´ÏìÖµ½øÐÐÕÉÁ¿£¬ÔÚƵƵÖظ´Ö®ºó£¬´«¸ÐÆ÷µÄ·´ÏìÊý¾ÝÖµ¿ÉÒÔ×é³ÉÒ»ÌõÖ±Ïߣ¬Õâ¸öÖ±ÏßÊÇÏßÐÔ±ä¸ïµÄ£¬´Ó¶ø¾Í¿ÉÒÔÖªµÀ´«¸ÐÆ÷ÊDz»ÊÇÓÐÎÊÌâÁË¡£

¡¡¡¡Because the tension testing machine system is a closed-loop system, in order to know which part of the system has problems, it is necessary to cut off the loop of the system and make it an open-loop system. So we can judge whether the sensor signal of the test machine is normal or not. At this time, the feedback signal of the displacement sensor should be disconnected to make it open-loop. At this time, the workstation sends out a signal to control the movement of the actuator, and measures the feedback value of the sensor. After several repetitions, the feedback data value of the sensor can form a straight line, which changes linearly, so that we can know whether the sensor is in trouble.

¡¡¡¡2¡¢Æä´Î¾ÍÊǶÔÀ­Á¦ÊÔÑé»úµÄ·Å´óÆ÷µ¥Î»µÄ¼ì²é£º

¡¡¡¡2. Next is the check of the amplifier unit of the tension testing machine.

¡¡¡¡ÔÚÒѾ­ÅųýÁËλÒÆ´«¸ÐÆ÷µÄÎÊÌâÖ®ºó£¬ÒòΪÀ­Á¦ÊÔÑé»úµÄϵͳÊÇÒ»¸ö±Õ»·ÏµÍ³£¬·Å´óÆ÷µ¥Î»µÄÕý³£Óë·ñÊÇÒ»¶¨Òª½øÐмì²éµÄ£¬ÒªÁì¾ÍÊǶϿª·Å´óÆ÷µ¥Î»µÄÊä³öÐźÅ¡£½«Êä³ö½ÓÏß²ð³ý£¬ÔÚ·Å´óÆ÷µ¥Î»¼ÓÉÏÒ»¸öÖ±Á÷ÊäÈëÐźÅ£¬È»ºóÕÉÁ¿°Ú´¸¹¥»÷ÊÔÑé»ú·Å´óÆ÷µ¥Î»µÄÊä³öÇé¿ö£¬ÔÚÖظ´ÊÔÑéƵƵ֮ºó£¬»ñµÃÒ»¸ö·Å´óÆ÷µ¥Î»Êä³öÐźŵÄÏßÐÔ±ä¸ïÇ÷ÊÆÊý¾Ý£¬ÕâÑù¾Í¿ÉÒÔÖªµÀ·Å´óÆ÷µ¥Î»ÊDz»ÊÇÓÐÎÊÌâÁË¡£

¡¡¡¡After eliminating the problem of displacement sensor, because the system of tension testing machine is a closed-loop system, whether the amplifier unit is normal or not must be checked by disconnecting the output signal of the amplifier unit. Remove the output wiring, add a DC input signal to the amplifier unit, and then measure the output of the amplifier unit of the pendulum impact tester. After repeated tests, the linear trend data of the output signal of an amplifier unit are obtained, so that we can know whether the amplifier unit is in trouble.

¡¡¡¡3¡¢×îºó¾ÍÊǼì²éÀ­Á¦ÊÔÑé»úµÄϵͳÉ趨ÊÇ·ñÕý³££º

¡¡¡¡3. Finally, check whether the system settings of the tension testing machine are normal:

¡¡¡¡ÊÂÇéÕ¾µ±ÖеÄÐж¯Æ÷µÄÐгÌÕý³£Çé¿öÏÂÓ¦¸ÃÊÇ¡À50mm£¬Èç¹ûÓëÕâ¸öÊý×Ö²»·û£¬ÄÇô¾ÍÊÇϵͳÉ趨µÄÎÊÌ⣬¾ÍÐèÒªÖØеÄÉ趨ϵͳ¡£ÒªÊÇÊÂÇéÕ¾µÄÏÔʾ°åÉÏÃæµÄÏÔʾֵÒÀ¾É²»ÕýÈ·£¬ÄÇô¾Í¿ÉÒÔÅж¨Ö»ÓÐÖØеı궨λÒÆϵͳ£¬²Å¿ÉÒÔ½â¾öÕâÒ»ÀàµÄÎÊÌâ¡£

¡¡¡¡Actuator travel in the workstation should normally be (+50mm). If it does not match this number, then it is the problem of system settings, and the system needs to be re-set. If the display value on the display board of the workstation is still incorrect, then it can be determined that only the re-positioning and moving system can solve this kind of problem.